Recalling Firm / Manufacturer |
St Jude Medical CRMD |
|
Product |
Implantable Cardioverter Defibrillator ATLAS II+ HF Tiered-therapy cardioverter/defibrillator with biventricular stimulation, Model V-367 |
|
Code Information |
All serial numbers |
|
Reason of Recall |
A condition that could lead to a ventricular sensing anomaly in Epic and Atlas family of lmplantable Cardioverter Defibrillators (ICDs) has been identified. A loss of ventricular sensing would prevent the ICD from being able to detect an arrhythmia. If an arrhythmia occurred and the device was unable to detect it, then a patient would not receive potentially lifesaving therapy. |
|
Date Recall Initiated |
January 16, 2008 |
|
Action |
A January 16, 2008, letter entitled, , sales representatives will follow-up with the physician and will upgrade all programmers.
In addition, St. Jude Medical issued a press release on January 18, 2008 and information concerning this advisory can be found on St. Jude's website, www.sirn.com. |
|
Contact |
Al-Jeel Medical & Trading Co. limited Riyadh (01) 4041717 |
|
Source of Information |
U.S.FDA |
|
SFDA Comments |
SFDA urges all hospitals that have devices subjected to recall, to contact the company. |