Hill-Rom Inc
24 صفر 1429 هـ الموافق 2 مارس 2008 م
|
Recalling Firm / Manufacturer |
Hill-Rom, Inc |
|
Product |
Hill-Rom Affinity Birthing Bed, Model P3700A |
|
Code Information |
All units. (**Affinity beds manufactured with serial number I290AA7395 or greater are not impacted by this correction.) |
|
Reason of Recall |
Hole/opening on the labor bar -- Amputation of a finger could result, if a finger is placed into the opening, as the hole/opening creates a shear point when the mechanism is activated |
|
Date Recall Initiated |
November 21, 2007 |
|
Action |
Consignees were notified of the problem via an Urgent Medical Device Correction letter dated 11/21/07. Users were informed that they would be receiving a free upgrade kit for all of their Affinity Birthing Beds. The Kit includes two plugs per bed, an insertion tool, and full written instructions. An instruction video is available at www.Hill-Rom.com |
|
Contact |
AMCO Riyadh (01) 4887122 |
|
Source of Information |
U.S. FDA |
|
SFDA Comments |
SFDA urges all hospitals that have devices subjected to recall, to contact the company. |