Philips Medical Systems


   29 شعبان 1428 هـ الموافق 11 سبتمبر 2007 م


Recalling Firm / Manufacturer

Philips Medical Systems

Product

Intellivue Central Station Software for the: Intellivue Information Center; Philips Information Center, Central Station Models: M3145, M3150, M3154, M3155, M3169

Code Information

Software versions: J.00.23 and J.00.24 (You can determine the application software release as follows: 1-Enter Service Mode through "All Controls" , "Service" 2-Under the "Other Services " menu item, click "Software Versions" 3-"Application" lists the application software version.

Reason of Recall

Incorrect Patient Assignment: the information center will in most cases be associated with the "incorrect" patient. This occurs on both standalone and networked systems running fixed mode monitoring. (This does not occur on systems that run flex mode monitoring)

Date Recall Initiated

July 16, 2007

Action

Philips Medical Systems issued a letter notification dated July 2007 " Urgent Device Correction". Customers with networked information centers will be corrected through a database server. Stand-Alone units will have a software correction. Software version J.00.25 will correct the problem. Users are advised not to use th 12 Lead Analysis and Export Option in fixed or flex mode monitoring unitl the software correction is made. Firm Initiated recall is ongoing.

Contact

Al-Faisaliah Medical Systems
Riyadh
(01) 4656551

Source of Information

U.S. FDA

SFDA Comments

SFDA urges all hospitals that have devices subjected to recall, to contact the company.