Philips Medical Systems
29 شعبان 1428 هـ الموافق 11 سبتمبر 2007 م
|
Recalling Firm / Manufacturer |
Philips Medical Systems |
|
Product |
Intellivue Central Station Software for the: Intellivue Information Center; Philips Information Center, Central Station Models: M3145, M3150, M3154, M3155, M3169 |
|
Code Information |
Software versions: J.00.23 and J.00.24 (You can determine the application software release as follows: 1-Enter Service Mode through "All Controls" , "Service" 2-Under the "Other Services " menu item, click "Software Versions" 3-"Application" lists the application software version. |
|
Reason of Recall |
Incorrect Patient Assignment: the information center will in most cases be associated with the "incorrect" patient. This occurs on both standalone and networked systems running fixed mode monitoring. (This does not occur on systems that run flex mode monitoring) |
|
Date Recall Initiated |
July 16, 2007 |
|
Action |
Philips Medical Systems issued a letter notification dated July 2007 " Urgent Device Correction". Customers with networked information centers will be corrected through a database server. Stand-Alone units will have a software correction. Software version J.00.25 will correct the problem. Users are advised not to use th 12 Lead Analysis and Export Option in fixed or flex mode monitoring unitl the software correction is made. Firm Initiated recall is ongoing. |
|
Contact |
Al-Faisaliah Medical Systems Riyadh (01) 4656551 |
|
Source of Information |
U.S. FDA |
|
SFDA Comments |
SFDA urges all hospitals that have devices subjected to recall, to contact the company. |