Abbott Laboratories


   23 شعبان 1428 هـ الموافق 5 سبتمبر 2007 م


Recalling Firm / Manufacturer

Abbott Laboratories

Product

Abbott CELL-DYN 3500 Hematology Analyzer

Code Information

Model number CELL-DYN 3500 CS, List Numbers 91340-01 and 91340-03, all serial numbers distributed before May 3, 2007. Model number CELL-DYN 3500 SL, List Numbers 91350-01 and 91350-03, all serial numbers distributed on or before May 3, 2007: 91350-01 1 EACH CD3500 SL ANALYZE(1 EACH) 1.000 0031192P, 91350-03 1 EACH CD3500SL ANAL-REF(1 EACH) 1.000 30336P96, 91350-03 1 EACH CD3500SL ANAL-REF 1.000 32159P96 05/05/2000, 91340-03 1 EACH CD 3300/3500CS(1 EACH) 1.000 30276N96 06/21/2000

Reason of Recall

The product failed electrostatic discharge testing, which could result in product failure.

Date Recall Initiated

May 11, 2007

Action

Mandatory Technical Service Bulletins (TSB) are being issued to Field Service to upgrade all instruments in the field. All customers will be contacted by Field Service to schedule this mandatory upgrade. The upgrades are expected to take 12 months. Firm Initiated recall is ongoing.

Contact

Abbott Diagnostics Division
Jeddah
(02) 6511077

Source of Information

U.S. FDA

SFDA Comments

SFDA urges all hospitals that have devices subjected to recall, to contact the company.