Summit Industries Inc.
26 شعبان 1428 هـ الموافق 8 سبتمبر 2007 م
|
Recalling Firm / Manufacturer |
Summit Industries Inc. |
|
Product |
LX125V Swivel Collimator, Model E600; Manual Radiographic Collimator; Summit Industries, Inc., Chicago, Illinois |
|
Code Information |
Model E600, serial numbers VA163-1192 to VB985-0798, all swivel units manufactured between 11/14/1992 and 07/31/1998 |
|
Reason of Recall |
The E600 Swivel Collimator may come loose from the tubestand and fall onto the x-ray table. |
|
Date Recall Initiated |
August 10, 2007 |
|
Action |
The affected dealers were notified of the field correction recall by letter dated 8/3/07, informing them of the problem with the stripped screws and requesting that they notify their affected end user locations. The end user notification letters inform them of the problem and state that the dealer will be contacting them soon to arrange a time to visit the end user location and conduct the rework on the collimator, which should take less than 30 minutes. In the mean time the end user was instructed to examine their swivel collimator for looseness. If it is found to be loose, the end user was instructed to take the unit out of service and call the dealer. Firm Initiated recall is ongoing. |
|
Contact |
N/A |
|
Source of Information |
U.S. FDA |
|
SFDA Comments |
SFDA urges all hospitals that have devices subjected to recall, to contact the company. |