Smiths Medical MD, Inc


   26 شعبان 1428 هـ الموافق 8 سبتمبر 2007 م


Recalling Firm / Manufacturer

Smiths Medical MD, Inc.

Product

Smiths CADD Yellow Medication Cassette Reservoir, With clamp and female Luer, Nonvented stopper included, 100 ml PC 4689-24A, REF 21-7100-24, Sterile, Made in Mexico Smith Medical MD, Inc. St Paul, MN 55112 USA

Code Information

083X16, 084X16, 116X16, 131X16,137X16, 140X16, 146X16, 151X16, 162X16, 185X16, 198X16, 224X16, 225X16

Reason of Recall

Smith Medical has become aware that an increase in pH may occur with some medications when instilled in the Affected Cassettes. Also, a small number of catheter occlusions have been reported from one customer when these Affected Cassettes have been used for delivering Flolan.

Date Recall Initiated

June 18, 2007

Action

All customers who have received the CADD Medication Cassette reservoirs have been sent the Product Recall Notification Letter on June 18, 2007 making them aware of the potential issue. Customers will be required to return the affected devices to Smiths Medical. All returned recall affected product will be quarantined and then destroyed. The consignees are to fill and fax "Attachment 1" form in regard to affected product inventory to Smiths Medical MD. Firm Initiated recall is ongoing.

Contact

Saudi Health Services (SHS)
Riyadh
(01) 4065912

Source of Information

U.S. FDA

SFDA Comments

SFDA urges all hospitals that have devices subjected to recall, to contact the company.