Medtronic MiniMed
04 رجب 1428 هـ الموافق 18 يوليو 2007 م
|
Recalling Firm / Manufacturer |
Medtronic MiniMed |
|
Product |
Medtronic MiniMed Paradigm Insulin Infusion Pumps for the following Paradigm models: MMT-511, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722 |
|
Code Information |
"All serial numbers" for the following Paradigm models: MMT-51 1, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722 |
|
Reason of Recall |
Exposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed Paradigm insulin infusion pump. Although there were alarms as a result of the damage, some users cleared these alarms and continued using the pump. Under such conditions, the pump will significantly over-deliver; potentially causing severe hypoglycemi |
|
Date Recall Initiated |
April 27, 2007 |
|
Action |
Firm mailed Medical Device Safety letters on April 24, 2007, to healthcare professionals, existing pump users and MRI facilities reiterating the existing warning in the pump User Guide to avoid exposing the pump to MRI (or similar high strength electromagnetic fields) and strengthening previous warning by specifically mentioning the potential for over-delivery and severe hypoglycemia. Firm is also including an insert with this information with any Paradigm infusion pumps shipped to new customers. Firm Initiated recall is ongoing. |
|
Contact |
Al-Jeel Medical & trading Co. ltd. 01 / 4041717 |
|
Source of Information |
U.S. FDA
|
|
SFDA Comments |
SFDA urges all hospitals that have devices subjected to recall, to contact the company. |