Medtronic MiniMed


   04 رجب 1428 هـ الموافق 18 يوليو 2007 م


Recalling Firm / Manufacturer

Medtronic MiniMed

Product

Medtronic MiniMed Paradigm Insulin Infusion Pumps for the following Paradigm models: MMT-511, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722

Code Information

"All serial numbers" for the following Paradigm models: MMT-51 1, MMT-512, MMT-712, MMT-515, MMT-715, MMT-522 and MMT-722

Reason of Recall

Exposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed Paradigm insulin infusion pump. Although there were alarms as a result of the damage, some users cleared these alarms and continued using the pump. Under such conditions, the pump will significantly over-deliver; potentially causing severe hypoglycemi

Date Recall Initiated

April 27, 2007

Action

Firm mailed Medical Device Safety letters on April 24, 2007, to healthcare professionals, existing pump users and MRI facilities reiterating the existing warning in the pump User Guide to avoid exposing the pump to MRI (or similar high strength electromagnetic fields) and strengthening previous warning by specifically mentioning the potential for over-delivery and severe hypoglycemia. Firm is also including an insert with this information with any Paradigm infusion pumps shipped to new customers. Firm Initiated recall is ongoing.

Contact

Al-Jeel Medical & trading Co. ltd.
01 / 4041717

Source of Information

U.S. FDA

 

SFDA Comments

SFDA urges all hospitals that have devices subjected to recall, to contact the company.