Abbott Diagnostic International, Ltd
08 رجب 1428 هـ الموافق 22 يوليو 2007 م
|
Recalling Firm / Manufacturer |
Abbott Diagnostic International, Ltd |
|
Product |
AxSYM Total B-hCG Reagent List 7A59-21 is composed of one bottle of Anti B-hCG Coated Microparticles, one bottle of Anti- B-hCG Alkaline Phosphatase Conjugate and one bottle of Specimen Diluent. REF 7A59-21. 100 Tests. For In Vitro Diagnostic Use. |
|
Code Information |
46329Q101, 47118Q101, 48063Q101, 48065Q101, 49341Q101, 49559Q101 |
|
Reason of Recall |
An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is magnified potentially yelding false positive results for pregnancy and or Dilution Inaccuracy. |
|
Date Recall Initiated |
April 10, 2007 |
|
Action |
The 4/10/2007 Product Recall Letter was sent to all customers who received the affected lots. Firm Initiated recall is ongoing. |
|
Contact |
Abbott Diagnostics Division Saudi Arabia (02) 6511077 |
|
Source of Information |
U.S. FDA |
|
SFDA Comments |
SFDA urges all hospitals that have devices subjected to recall, to contact the company. |