Abbott Diagnostic International, Ltd


   08 رجب 1428 هـ الموافق 22 يوليو 2007 م


Recalling Firm / Manufacturer

Abbott Diagnostic International, Ltd

Product

Total B-hCG Speciment Diluent is composed of 1 bottle of Total B-hCG Specimen Diluent containing animal serum.

Code Information

45033Q100, 46212Q100,46776Q100, 47117Q100, 47619Q100

Reason of Recall

An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is magnified potentially yelding false positive results for pregnancy and or Dilution Inaccuracy

Date Recall Initiated

April 10, 2007

Action

The 4/10/2007 Product Recall Letter was sent to all customers who received the affected lots. Firm Initiated recall is ongoing

Contact

Abbott Diagnostics Division
Saudi Arabia
(02) 6511077

Source of Information

U.S. FDA

SFDA Comments

SFDA urges all hospitals that have devices subjected to recall, to contact the company.