Recalling Firm / Manufacturer |
Teleflex Creek Dr |
|
Product |
Various Sizes of Sterile Polygycolic Acid Synthetic Surgical Suture- Bondek?, TOPS ASSY P/P SLING, Catalog #84414323 |
|
Code Information |
Bondek®, TOPS ASSY P/P SLING, Catalog #84414323, Lot #9591. |
|
Reason of Recall |
Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage. |
|
Date Recall Initiated |
June 06, 2007 |
|
Action |
Consignees were notified by letter on/about 06/07/2007. They were advised to cease use, remove and quarantine all affected product immediately. Affected products are to be returned to Teleflex Medical. An Acknowledgement & Stock form was included to be completed and returned via fax to Teleflex Medical. Firm Initiated recall is ongoing. |
|
Contact |
AMHSCO Riyadh (01) 4621188 |
|
Source of Information |
U.S FDA |
|
SFDA Comments |
SFDA urges all hospitals that have devices subjected to recall, to contact the company. |